This CAPA addresses a flow rate over-delivery failure identified during final QC inspection of the FlowMaster IV Infusion Pump (FMP-300), LOT-2026-0041. Testing at the 125 mL/hr setting revealed flow rates 14.8% above setpoint — 2.96× the ±5% specification tolerance. 23 units from this lot are affected. No patient involvement has been confirmed; all units were intercepted pre-shipment.
The CRITICAL classification is driven by three factors acting together: this is a life-sustaining device (infusion pump), the failure mode directly affects primary device function (drug delivery accuracy), and the magnitude of the deviation (14.8% vs. ±5% tolerance) is substantial. Sterility or life-sustaining device failures are never classified Minor under ISO 13485 §8.5.2 regardless of detection point or patient involvement. The CRITICAL tier carries a 30-day root cause requirement and 90-day closure target.
Field action risk is assessed as HIGH. While all known affected units were intercepted in-house, the first step in the investigation must confirm that no units from this lot or adjacent lots were distributed. If distribution is confirmed, the flow rate error profile of this device class — used for IV medication delivery including drugs with narrow therapeutic windows — creates potential for patient harm. A field safety evaluation under 21 CFR Part 806 would be required in that scenario.
Root cause analysis should focus on the flow sensor calibration chain: sensor supplier calibration certificate, incoming inspection acceptance criteria, and in-process calibration verification during assembly. The IS / IS NOT analysis constrains scope to this lot and this failure mode — expanding scope prematurely to other models or settings before root cause is established is a common CAPA failure pattern that extends timelines without improving outcomes.
| Tool | Regulation / Standard | Decision Logic | Output Type |
|---|---|---|---|
| Problem Statement | IS / IS NOT Method | Formal statement + IS / IS NOT matrix from raw inputs — prevents vague scope at opening | Generated |
| Priority & Classification | ISO 13485:2016 §8.5.2 | CRITICAL / MAJOR / MINOR / OBSERVATION; life-sustaining and sterility flags override to CRITICAL/MAJOR | Automated |
| Scope Boundary | ISO 13485:2016 §8.5.2(a) | In scope / out of scope / open questions; limits scope creep while ensuring full coverage | Generated |
| Containment Plan | ISO 13485:2016 §8.5.3 | Immediate actions (today) vs. short-term (48 hrs) — urgency-flagged and role-assigned | Generated |
| Field Action Risk | 21 CFR Part 806 · ISO 14971 | HIGH / MEDIUM / LOW based on patient exposure, device class, and failure mode severity | Automated |
| Deadline Calculation | ISO 13485:2016 §8.5.2 | Containment, root cause, actions, closure — calculated from date opened and priority tier | Calculated |
| Investigation Agenda | CAPA SOP / ISO 13485 | Prioritized steps with task description, owner, and specific regulatory reference for each action | Generated |